5-144
IM 701310-17E
:TELecomtest:EYEPattern:TLEVels:
PERCent
Function Sets the threshold level of the eye pattern as a
percentage or queries the current setting.
Syntax
:TELecomtest:EYEPattern:TLEVels:
PERCent {<NRf>,<NRf>}
:TELecomtest:EYEPattern:TLEVels:
PERCent?
<NRf> = 0 to 100 (%)
Example
:TELECOMTEST:EYEPATTERN:TLEVELS:
PERCENT 90,10
:TELECOMTEST:EYEPATTERN:TLEVELS:
PERCENT? -> :TELECOMTEST:EYEPATTERN:
TLEVELS:PERCENT 90,10
:TELecomtest:EYEPattern:TLEVels:UNIT
Function Sets the threshold level of the eye pattern in UNIT or
queries the current setting.
Syntax
:TELecomtest:EYEPattern:TLEVels:
UNIT {<NRf>,<NRf>|<Voltage>,<Voltage>|
<Current>,<Current>}
:TELecomtest:EYEPattern:TLEVels:UNIT?
<NRf>, <Voltage>, and <Current> = See the DL9000
User
’s Manual.
Example
:TELECOMTEST:EYEPATTERN:TLEVELS:
UNIT 1,0
:TELECOMTEST:EYEPATTERN:TLEVELS:UNIT?
-> :TELECOMTEST:EYEPATTERN:TLEVELS:
UNIT 1.000E+00,0.000E+00
:TELecomtest:EYEPattern:VDARk
Function Sets the dark level (zero light level) or queries the
current setting.
Syntax
:TELecomtest:EYEPattern:VDARk {<NRf>|
<Voltage>|<Current>}
:TELecomtest:EYEPattern:VDARk?
<NRf>, <Voltage>, and <Current> = See the DL9000
User
’s Manual.
Example
:TELECOMTEST:EYEPATTERN:VDARK 1.000E+00
:TELECOMTEST:EYEPATTERN:VDARK?
-> :TELECOMTEST:EYEPATTERN:
VDARK 1.000E+00
:TELecomtest:EYEPattern:<Parameter>:
STATe
Function Turns ON/OFF the waveform parameter of the eye
pattern or queries the current setting.
Syntax
:TELecomtest:EYEPattern:<Parameter>:
STATe {<Boolean>}
:TELecomtest:EYEPattern:<Parameter>:
STATe?
<Parameter> = {DBERate|EHEight|EWIDth|
FALL|JITT
er<x>|PCROssing|PDUTycycle|
QFACtor|RISE|SDBase|SDTop|T1CRossing|
T2CRossing|VBASe|VCRossing|VTOP}
<x> = 1 or 2
Example (The following is an example for DBERate.)
:TELECOMTEST:EYEPATTERN:DBERATE:
STATE ON
:TELECOMTEST:EYEPATTERN:DBERATE:STATE?
-> :TELECOMTEST:EYEPATTERN:DBERATE:
STATE 1
:TELecomtest:EYEPattern:<Parameter>:
VALue?
Function Queries the waveform parameter value of the eye
pattern.
Syntax
:TELecomtest:EYEPattern:<Parameter>:
VALue?
<Parameter> = {DBERate|EHEight|EWIDth|
FALL|JITT
er<x>|PCROssing|PDUTycycle|
QFACtor|RISE|SDBase|SDTop|T1CRossing|
T2CRossing|VBASe|VCRossing|VTOP}
<x> = 1 or 2
Example (The following is an example for DBERate.)
:TELECOMTEST:EYEPATTERN:DBERATE:VALUE?
-> :TELECOMTEST:EYEPATTERN:DBERATE:
VALUE 1.000E+00
:TELecomtest:EYEPattern:TLEVels?
Function Queries all settings related to the threshold level of
the eye pattern.
Syntax
:TELecomtest:EYEPattern:TLEVels?
Example
:TELECOMTEST:EYEPATTERN:TLEVELS?
-> :TELECOMTEST:EYEPATTERN:TLEVELS:
MODE PERCENT;PERCENT 90,10;
UNIT 1.000E+00,0.000E+00
:TELecomtest:EYEPattern:TLEVels:MODE
Function Sets the unit of the threshold level of the eye pattern
or queries the current setting.
Syntax
:TELecomtest:EYEPattern:TLEVels:
MODE {PERCent|UNIT}
:TELecomtest:EYEPattern:TLEVels:MODE?
Example
:TELECOMTEST:EYEPATTERN:TLEVELS:
MODE PERCENT
:TELECOMTEST:EYEPATTERN:TLEVELS:MODE?
-> :TELECOMTEST:EYEPATTERN:TLEVELS:
MODE PERCENT
5.27 TELecomtest Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals