EasyManua.ls Logo

YOKOGAWA DL9040 - Page 215

YOKOGAWA DL9040
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-181
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:PATTern
Function Sets each data of the OR condition of the CAN bus
signal trigger in binary notation or queries the current
setting.
Syntax
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:PATTern {<String>}
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:PATTern?
<x> = 1 to 4
<String> = Up to 64 characters by combining ‘0,’
‘1,’
and ‘X’ (in one-byte unit)
Example
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
DATA:PATTERN "11011111"
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
DATA:PATTERN? -> :TRIGGER:ENHANCED:
CANBUS:IDOR:ID1:DATA:PATTERN
"11011111"
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:SIGN
Function Sets sign of each data of the OR condition of the
CAN bus signal trigger or queries the current setting.
Syntax
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:SIGN {SIGN|UNSign}
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:SIGN?
<x> = 1 to 4
Example
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
DATA:SIGN SIGN
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
DATA:SIGN? -> :TRIGGER:ENHANCED:
CANBUS:IDOR:ID1:DATA:SIGN SIGN
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
FORMat
Function Sets each message format (standard or extended)
of the OR condition of the CAN bus signal trigger or
queries the current setting.
Syntax
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
FORMat {STD|EXT}
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
FORMat?
<x> = 1 to 4
Example
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
FORMAT STD
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
FORMAT? -> :TRIGGER:ENHANCED:CANBUS:
IDOR:ID1:FORMAT STD
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:DLC
Function Sets the number of valid bytes (DLC) of each data
of the OR condition of the CAN bus signal trigger or
queries the current setting.
Syntax
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:DLC {<NRf>}
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:DLC?
<x> = 1 to 4
<NRf> = 0 to 8
Example
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
DATA:DLC 0
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
DATA:DLC? -> :TRIGGER:ENHANCED:
CANBUS:IDOR:ID1:DATA:DLC 0
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:HEXA
Function Sets each data of the OR condition of the CAN bus
signal trigger in hexadecimal notation.
Syntax
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:HEXA {<String>}
<x> = 1 to 4
<String> = Up to 16 characters by combining ‘0’
to ‘F’
and ‘X’ (in one-byte unit)
Example
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
DATA:HEXA "A9"
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:MSBLsb
Function Sets the MSB and LSB bits of each data of the OR
condition of the CAN bus signal trigger or queries the
current setting.
Syntax
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:MSBLsb {<NRf>,<NRf>}
:TRIGger:ENHanced:CANBus:IDOR:ID<x>:
DATA:MSBLsb?
<x> = 1 to 4
<NRf> = See the User's Manual (IM701310-01E).
Example
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
DATA:MSBLSB 1,0
:TRIGGER:ENHANCED:CANBUS:IDOR:ID1:
DATA:MSBLSB? -> :TRIGGER:ENHANCED:
CANBUS:IDOR:ID1:DATA:MSBLSB 1,0
5.29 TRIGger Group

Table of Contents

Related product manuals