5-42
IM 701310-17E
:ANALysis:AHIStogram<x>:MEASure:
PARameter:HRANge<x>
Function Sets the horizontal range of the waveform parameter
of the accumulated histogram or queries the current
setting.
Syntax
:ANALysis:AHIStogram<x>:MEASure:
PARameter:HRANge<x> {<NRf>,<NRf>}
:ANALysis:AHIStogram<x>:MEASure:
PARameter:HRANge<x>?
<x> of AHIStogram<x> = 1 or 2
<x> of HRANge<x> = 1 or 2
<NRf> = –5 to 5 div
Example
:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:HRANGE1 1,2
:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:HRANGE1? -> :ANALYSIS:
AHISTOGRAM1:MEASURE:
PARAMETER:HRANGE1 2.000E+00,1.000E+00
:ANALysis:AHIStogram<x>:MEASure:
PARameter:VRANge<x>
Function Sets the vertical range of the waveform parameter
of the accumulated histogram or queries the current
setting.
Syntax
:ANALysis:AHIStogram<x>:MEASure:
PARameter:VRANge<x> {<NRf>,<NRf>}
:ANALysis:AHIStogram<x>:MEASure:
PARameter:VRANge<x>?
<x> of AHIStogram<x> = 1 or 2
<x> of VRANge<x> = 1 or 2
<NRf> = –4 to 4 (div)
Example
:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:VRANGE1 1,2
:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:VRANGE1? -> :ANALYSIS:
AHISTOGRAM1:MEASURE:
PARAMETER:VRANGE1 2.000E+00,1.000E+00
:ANALysis:AHIStogram<x>:MODE
Function Sets the accumulated histogram mode or queries the
current setting.
Syntax
:ANALysis:AHIStogram<x>:MODE
{HORizontal|VERTical}
:ANALysis:AHIStogram<x>:MODE?
<x> = 1 or 2
Example
:ANALYSIS:AHISTOGRAM1:MODE HORIZONTAL
:ANALYSIS:AHISTOGRAM1:MODE?
->
:ANALYSIS:AHISTOGRAM1:MODE HORIZONTAL
:ANALysis:AHIStogram<x>:TRACe
Function Sets the source trace of the accumulated histogram
or queries the current setting.
Syntax
:ANALysis:AHIStogram<x>:TRACe {<NRf>}
:ANALysis:AHIStogram<x>:TRACe?
<x> = 1 or 2
<NRf> = 1 to 8
Example
:ANALYSIS:AHISTOGRAM1:TRACE 1
:ANALYSIS:AHISTOGRAM1:TRACE?
-> :ANALYSIS:AHISTOGRAM1:TRACE 1
:ANALysis:AHIStogram<x>:VERTical
Function Sets the vertical range of the accumulated histogram
or queries the current setting.
Syntax
:ANALysis:AHIStogram<x>:VERTical
{<NRf>,<NRf>}
:ANALysis:AHIStogram<x>:VERTical?
<x> = 1 or 2
<NRf> = –4 to 4 (div)
Example
:ANALYSIS:AHISTOGRAM1:VERTICAL 1,2
:ANALYSIS:AHISTOGRAM1:VERTICAL?
-> :ANALYSIS:AHISTOGRAM1:
VERTICAL 2.000E+00,1.000E+00
:ANALysis:AHIStogram<x>:WINDow
Function Sets the measurement target window of the
accumulated histogram or queries the current setting.
Syntax
:ANALysis:AHIStogram<x>:WINDow {MAIN|
Z1|Z2}
:ANALysis:AHIStogram<x>:WINDow?
<x> = 1 or 2
Example
:ANALYSIS:AHISTOGRAM1:WINDOW MAIN
:ANALYSIS:AHISTOGRAM1:WINDOW?
-> :ANALYSIS:AHISTOGRAM1:WINDOW MAIN
:ANALysis:DISPlay<x>
Function Turns ON/OFF the analysis function display or
queries the current setting.
Syntax
:ANALysis:DISPlay<x> {<Boolean>}
:ANALysis:DISPlay<x>?
<x> = 1 or 2
Example
:ANALYSIS:DISPLAY1 ON
:ANALYSIS:DISPLAY1? -> :ANALYSIS:
DISPLAY1 1
5.3 ANALysis Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals