EasyManua.ls Logo

YOKOGAWA DL9040

YOKOGAWA DL9040
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-42
IM 701310-17E
:ANALysis:AHIStogram<x>:MEASure:
PARameter:HRANge<x>
Function Sets the horizontal range of the waveform parameter
of the accumulated histogram or queries the current
setting.
Syntax
:ANALysis:AHIStogram<x>:MEASure:
PARameter:HRANge<x> {<NRf>,<NRf>}
:ANALysis:AHIStogram<x>:MEASure:
PARameter:HRANge<x>?
<x> of AHIStogram<x> = 1 or 2
<x> of HRANge<x> = 1 or 2
<NRf> = –5 to 5 div
Example
:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:HRANGE1 1,2
:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:HRANGE1? -> :ANALYSIS:
AHISTOGRAM1:MEASURE:
PARAMETER:HRANGE1 2.000E+00,1.000E+00
:ANALysis:AHIStogram<x>:MEASure:
PARameter:VRANge<x>
Function Sets the vertical range of the waveform parameter
of the accumulated histogram or queries the current
setting.
Syntax
:ANALysis:AHIStogram<x>:MEASure:
PARameter:VRANge<x> {<NRf>,<NRf>}
:ANALysis:AHIStogram<x>:MEASure:
PARameter:VRANge<x>?
<x> of AHIStogram<x> = 1 or 2
<x> of VRANge<x> = 1 or 2
<NRf> = –4 to 4 (div)
Example
:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:VRANGE1 1,2
:ANALYSIS:AHISTOGRAM1:MEASURE:
PARAMETER:VRANGE1? -> :ANALYSIS:
AHISTOGRAM1:MEASURE:
PARAMETER:VRANGE1 2.000E+00,1.000E+00
:ANALysis:AHIStogram<x>:MODE
Function Sets the accumulated histogram mode or queries the
current setting.
Syntax
:ANALysis:AHIStogram<x>:MODE
{HORizontal|VERTical}
:ANALysis:AHIStogram<x>:MODE?
<x> = 1 or 2
Example
:ANALYSIS:AHISTOGRAM1:MODE HORIZONTAL
:ANALYSIS:AHISTOGRAM1:MODE?
->
:ANALYSIS:AHISTOGRAM1:MODE HORIZONTAL
:ANALysis:AHIStogram<x>:TRACe
Function Sets the source trace of the accumulated histogram
or queries the current setting.
Syntax
:ANALysis:AHIStogram<x>:TRACe {<NRf>}
:ANALysis:AHIStogram<x>:TRACe?
<x> = 1 or 2
<NRf> = 1 to 8
Example
:ANALYSIS:AHISTOGRAM1:TRACE 1
:ANALYSIS:AHISTOGRAM1:TRACE?
-> :ANALYSIS:AHISTOGRAM1:TRACE 1
:ANALysis:AHIStogram<x>:VERTical
Function Sets the vertical range of the accumulated histogram
or queries the current setting.
Syntax
:ANALysis:AHIStogram<x>:VERTical
{<NRf>,<NRf>}
:ANALysis:AHIStogram<x>:VERTical?
<x> = 1 or 2
<NRf> = –4 to 4 (div)
Example
:ANALYSIS:AHISTOGRAM1:VERTICAL 1,2
:ANALYSIS:AHISTOGRAM1:VERTICAL?
-> :ANALYSIS:AHISTOGRAM1:
VERTICAL 2.000E+00,1.000E+00
:ANALysis:AHIStogram<x>:WINDow
Function Sets the measurement target window of the
accumulated histogram or queries the current setting.
Syntax
:ANALysis:AHIStogram<x>:WINDow {MAIN|
Z1|Z2}
:ANALysis:AHIStogram<x>:WINDow?
<x> = 1 or 2
Example
:ANALYSIS:AHISTOGRAM1:WINDOW MAIN
:ANALYSIS:AHISTOGRAM1:WINDOW?
-> :ANALYSIS:AHISTOGRAM1:WINDOW MAIN
:ANALysis:DISPlay<x>
Function Turns ON/OFF the analysis function display or
queries the current setting.
Syntax
:ANALysis:DISPlay<x> {<Boolean>}
:ANALysis:DISPlay<x>?
<x> = 1 or 2
Example
:ANALYSIS:DISPLAY1 ON
:ANALYSIS:DISPLAY1? -> :ANALYSIS:
DISPLAY1 1
5.3 ANALysis Group

Table of Contents

Related product manuals