Loading...
2-10
IM 701310-01E
Pulse State
A trigger is activated in any of the following cases.
When the time during which the
state condition is met or not met satisfies the
relationship with the specified determination time.
The DL9000 checks and normalizes the state condition on the
rising or falling edge
of the specified signal (clock source). A
trigger is activated when the time during
which the normalized condition is met or not met first satisfies the relationship with
the specified time.
L
H
H
CH1
CH2
State: CH1 = H, CH2 = L, CH3/CH4 = X, AND
Clock: None, Polarity: True, More than
L: low level, H: high level
Trigger
CH1
CH2
H
H
L
L
L
L
H
H
CH1
CH2
CH1
CH2
H
H
L
L
L
H
H
CH1
CH2
State: CH1 = H, CH2 = L, CH3/CH4 = X, AND
Clock: None, Polarity: False, More than
L: low level, H: high level
Trigger
CH1
CH2
L
H
L
State: CH1 = H, CH2 = L, CH4 = X, AND
Clock: CH3, rising, Polarity: True, More than
L: low level, H: high level
State condition
t
T1
t > T1
L
State condition
t
T1
t > T1
t
T1
t > T1
State: CH1 = H, CH2 = L, CH4 = X, AND
Clock: CH3, rising, Polarity: False, More than
L: low level, H: high level
CH1
CH2
H
H
CH1
CH2
L
H
L
L
Trigger
CH3
Normalized condition
t
T1
t > T1
Trigger
CH3
Normalized condition
State condition false
State condition false
State condition true
State condition false State condition false
State condition true
State condition false State condition false
State condition true
State condition true State condition true
State condition
false