6-7
IM DLM5058-03EN
Specifications
1
2
3
4
5
6
App
Index
Item Specifications
PSI5 Airbag
1
: Triggers on PSI5 Airbag signal
The source can be selected from CH1 to CH8.
2
Mode: Sync, Start Bit, Frame in Slot, Data, Error
Bit Rate: 125kbps, 189kbps, or User Define
For User Define, you can set a value from 10.0 kbps to 1000.0 kbps in 0.1-
kbps steps.
UART
1
: Triggers on a UART (RS232) signal
The source can be selected from CH1 to CH8,
2
and Logic.
Mode: Every Data, Error, Data
Format: 8bit No Parity, 7bit Parity, 8bit Parity
Bit Rate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200bps
or User Define
For User Define, you can set a value from 1 k to 10 Mbps in 0.5-bps steps.
I
2
C
1
: Triggers on a I
2
C bus signal
The source can be selected from CH1 to CH8,
2
and Logic.
Mode: Every Start, Address Data, NON ACK, General Call, Start Byte,
HS Mode
SPI
1
: Triggers on an SPI (Serial Peripheral Interface) bus signal
The source can be selected from CH1 to CH8,
2
and Logic.
Mode: 3wire, 4wire
User Define: Triggers on a general-purpose serial communication signal
The source can be selected from CH1 to CH8.
2
Data source, enable source, clock source, and latch source can be specified.
Bitrate: 1 k to 200 Mbps (with clock)
1 k to 50 Mbps (without clock)
Bit length: 1 bits to 128 bits
TV: Triggers on the specified field number, line number, or polarity in video signals of
various broadcasting formats.
The source can be selected from CH1 to CH8.
2
Mode: NTSC: Triggers on an NTSC (525/60/2) signal
PAL: Triggers on an PAL (625/50/2) signal
SDTV: Triggers on an SDTV (480/60p) signal
HDTV: Triggers on the following HDTV signals
1080/60i, 1080/50i, 720/60p, 1080/25p,
1080/24p, 1080/24sF, 1080/60p
UserdefTV: You can trigger on any TV signal by selecting standard or high
definition, setting the Hsync period, and setting the sync guard.
Sync guard can be set to a value from 60 to 90% of the Hsync
value in 1% steps.
Porality: Pos, Neg
HF Rej: NTSC/PAL: 300 kHz (fixed)
SDTV/HDTV: OFF (fixed)
UserdefTV: Off or 300 kHz
Line: 5 to 1054 (NTSC), 2 to 1251 (PAL), 8 to 2251 (SDTV), 2 to 2251 (HDTV),
2 to 2251 (UserdefTV), ALL
Field: 1, 2, X
Frame Skip: 1, 2, 4, 8
AB trigger Triggers on the combined conditions of the B trigger conditions and A trigger conditions
Some trigger types cannot be specified depending on the AB trigger type shown below.
The specifications of the types that trigger B can be set to are the same as those of trigger A.
Serial bus trigger can only be used on either trigger A or trigger B.
(See “Trigger type (A trigger)” in section 6.2, “Triggering Section.”)
OFF: Triggers only on the trigger A conditions (the trigger B conditions are not used).
A Delay B: After the trigger A conditions are met and the specified amount of time elapses, the
instrument triggers when the trigger B conditions are met.
Trigger B cannot be set to Pattern (with time conditions), Pulse Width, Rise/Fall
Time, RUNT, Timeout, Window (with time conditions), or TV.
Time setting: 10 ns to 10 s
A to B (N): After the trigger A conditions are met, the instrument triggers when the trigger B
conditions are met N times.
Trigger B cannot be set to Pattern (with time conditions), Pulse Width, Rise/Fall
Time, RUNT, Timeout, Window (with time conditions), or TV.
Count setting: 1 to 10
9
Forced trigger A trigger that is forced, regardless of whether the trigger conditions are met.
1 FlexRay, CAN, CAN FD, LIN, CXPI, SENT, PSI5 Airbag, UART, I2C, and SPI are options.
2 CH1 to CH4 for the DLM5034, DLM5054
6.2 Triggering Section