When the Set Record Length Is 500 Mpoint
(Selectable on models with the /M2S or /M2 option)
Mode
Setting
High Resolution Mode OFF High Resolution Mode ON
RealTime Interpolation Repetitive RealTime Interpolation Repetitive
Time/div
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
500 s
100 k 500 M
Interpolation mode and Repetitive
sampling mode cannot be used
with this record length.
125 k 500 M
Interpolation mode and Repetitive
sampling mode cannot be used
with this record length.
200 s
250 k 500 M 250 k 500 M
100 s
500 k 500 M 500 k 500 M
50 s
1 M 500 M 1 M 500 M
20 s
2.5 M 500 M 2.5 M 500 M
10 s
5 M 500 M 5 M 500 M
5 s
10 M 500 M 10 M 500 M
2 s
25 M 500 M 25 M 500 M
1 s
50 M 500 M 50 M 500 M
500 ms
100 M 500 M 100 M 500 M
200 ms
250 M 500 M 250 M 500 M
100 ms
500 M 500 M 500 M 500 M
50 ms
625 M 312.5 M 625 M 312.5 M
20 ms
2.5 G 500 M 1.25 G 250 M
10 ms
2.5 G 250 M 1.25 G 125 M
5 ms
2.5 G 125 M 1.25 G 50 M
2 ms
2.5 G 50 M 1.25 G 25 M
1 ms
2.5 G 25 M 1.25 G 12.5 M
500 µs
2.5 G 12.5 M 1.25 G 6.25 M
200 µs
2.5 G 5 M 1.25 G 2.5 M
100 µs
2.5 G 2.5 M 1.25 G 1.25 M
50 µs
2.5 G 1.25 M 1.25 G 625 k
20 µs
2.5 G 500 k 1.25 G 250 k
10 µs
2.5 G 250 k 1.25 G 125 k
5 µs
2.5 G 125 k 1.25 G 62.5 k
2 µs
2.5 G 50 k 1.25 G 25 k
1 µs
2.5 G 25 k 1.25 G 12.5 k
500 ns
2.5 G 12.5 k 1.25 G 6.25 k
200 ns
2.5 G 5 k 1.25 G 2.5 k
100 ns
2.5 G 2.5 k 1.25 G 1.25 k
50 ns
2.5 G 1.25 k 1.25 G 625
20 ns
2.5 G 500 1.25 G 250
10 ns
2.5 G 250 1.25 G 125
5 ns
2.5 G 125 1.25 G 62.5
2 ns
2.5 G 50 1.25 G 25
1 ns
2.5 G 25 1.25 G 12.5
Acquisition when you specify this record length is performed in single mode.
: Roll mode
Appendix 1 Relationship between the Time Axis Setting, Sample Rate, and Record Length

Table of Contents

Other manuals for YOKOGAWA DLM5034

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM5034 and is the answer not in the manual?

YOKOGAWA DLM5034 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM5034
CategoryTest Equipment
LanguageEnglish

Related product manuals