EasyManuals Logo

YOKOGAWA WT1803 User Manual

YOKOGAWA WT1803
167 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #85 background imageLoading...
Page #85 background image
9-10
IM WT1801-01EN
9 Integrated Power (Watt hours)
Integration Timer (Integ Timer)
You can set the hour, minute, and second within the following range.
00000:00:00 to 10000:00:00
When Independent Integration Is Disabled
The integration timer that you specified above applies to every element.
When Independent Integration Is Enabled
Setting
Y
ou can select the method for setting the integration timer from the following options.
Each: Set it separately for each input element.
All: Set it for all installed input elements at the same time.
Element1 to Element6
You can set the integration timers for each element within the range given above.
Scheduled Times for Real-Time Integration (Real-time Control)
This setting is valid in real-time normal integration mode and real-time continuous integration mode.
You can set the year, month, day, hour, minute, and second of the integration start and stop times. Be sure to
set the integration stop time to a time after the integration start time. You can set the values within the following
ranges.
Y
ear: Any four-digit Gregorian calendar value
Hour:Minute:Second: 00:00:00 to 23:59:59
Now: The scheduled integration start time is set to the current time.
Copy: The scheduled integration start time is copied to the scheduled integration stop time.
When Independent Integration Is Disabled
The scheduled times that you specified above apply to every element.
When Independent Integration Is Enabled
Setting
Y
ou can select the method for setting the scheduled times from the following options.
Each: Set them separately for each input element.
All: Set them for all installed input elements at the same time.
Element1 to Element6
You can set the scheduled times for each element within the range given above.
You can set the day for the scheduled start or stop time in February to a value as high as the 31st day. If
you do so, an error message will appear when you start integration. Reset the scheduled stop time.
The WT1800 recognizes leap years when it executes the integ
ration operation.
In real-time normal integration mode and real-time continuous integration mode, after you press Start and
the WT1800 enters into an integration-ready state, the numeric data may not be updated immediately.
This is because the numeric data is updated in sync with the WT1800 clock. This ensures that the numeric
data is updated at the same time as the scheduled integration start time and that the integration time is
accurate.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA WT1803 and is the answer not in the manual?

YOKOGAWA WT1803 Specifications

General IconGeneral
BrandYOKOGAWA
ModelWT1803
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals