Automated night blood pressure
ANBP
The ANBP can specify up to six preset start time a day.
When parameters for the ANBP program are stored, the ANBP is
started and a "1 scope" is performed for each start time.
"1 scope" consists of a set of blood pressure measurements and
interval times which are repeated to the number of the
measurement count. The last interval time is omitted.
Note
Manual blood pressure measurement of the ANBP
When the / switch is pressed in S-BPM waiting mode, a "1
scope" is performed.
T : Interval time
1 scope
T
Time
T T
T
The ANBP is started at the same time when ANBP program is set.
(The ANBP becomes waiting mode until start time)
Start time
Start time
Time
T T T T
Press the
/ switch
to start the manual-BPM.
1 scope
T
T
1 scope
1 scope
1 scope
Time
T T
T
T
Press the
/ switch
to stop the "1 scope".
Not measured
1 scope
1 scope
Blood pressure measurement
Start time
Start time
Start time