102 | NGC8200 AND PGC1000 USER MANUAL | 2101510MNAH
To update:
1. Click the 32 Bit loader icon on the PCCU top menu.
Figure 10-4: Select 32-bit loader
2. Click Yes at the warning if you have backed up your device data. If you have not backed up the
analyzer, see section 10.6 Back up configuration files.
3. At the Connection Setup window (Figure 10-5):
a. Select Network in the Connection Type drop-down list. "Network" is the correct option,
even though the physical Ethernet is point-to-point. A point-to-point connection is
recommended for update.
b. Type the IP address of the target analyzer in the IP Address field.
Figure 10-5: Connection Setup
4. Click Connect. The loader screen displays.
IMPORTANT NOTE: Always verify that in the Connection field, at the bottom left corner of the
loader screen, the option: Network displays; and that the IP address is the address of the target
analyzer. Avoid upgrading the wrong device. Direct connection to the device is always
recommended to avoid connecting to the wrong device over a network.
5. Load the Flash (Figure 10-6):
a. Click the Flash browse button to locate the Flash file in your system.
b. Select the Flash file: 2102411-0xxEX.ngc. Where xx is the version of the update package.