FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D 131
13 Specification
Note
The device data sheet is available in the ABB download area at
www.abb.com/flow.
Permitted pipe vibration
In accordance with EN 60068-2-6
Valid for sensors in remote mount and integral mount design.
Maximum deflection: 0.15 mm (0.006 in) in the frequency
range of 10 to 58 Hz
Maximum acceleration: 2 g in the frequency range of
58 to 150 Hz
ProcessMaster – Temperature data
The temperature range offered by the device is dependent on a
number of different factors.
These factors include the measuring medium temperature
T
medium
, the ambient temperature T
amb
, operating pressure
P
medium
, liner material and the approval for explosion protection.
Storage temperature range
40 to 70 °C (40 to 158 °F)
Maximum permissible cleaning temperature
CIP media Liner
Cleaning temperature
Steam PTFE, PFA
150 °C (302 °F)
Cleaning fluid PTFE, PFA
140 °C (284 °F)
The maximum cleaning temperature specified applies to a
maximum ambient temperature of 25 °C (77 °F).
If the ambient temperature up-scales > 25 °C (> 77 °F), then
the temperature difference to the current temperature must
be subtracted from the max. cleaning temperature.
The specified cleaning temperature may have an effect for a
maximum of 60 minutes.
Change from two to one column

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB ProcessMaster FEP630 series and is the answer not in the manual?

ABB ProcessMaster FEP630 series Specifications

General IconGeneral
BrandABB
ModelProcessMaster FEP630 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals