2105552MNAE | RMC-100 | 135
Figure 7-13: High calibration value reading
22. Adjust the calibrator to apply the 50% value of the external device signal range. For example, if
the signal range of the external device is 1-5 Vdc, apply 3 Vdc.
23. Click 50% Cal Point. The Enter 50% Calibration Value dialog box displays with the target mid-
value automatically calculated. In Figure 7-14
, the box displays 750.000. This is the mid-value
of 1500.
24. Wait for the value in the Current Reading field to stabilize. This value should match or be close
to the signal value from the calibrator (~ 3 Vdc).
Figure 7-14: User-defined mid calibration value
25. Click OK. The Calibration Complete message displays (Figure 7-15
).