QUALITY CONTROLS - 5
VerifyNow System User Manual Page 27 14340.J
Figure 5-1 Main Menu
A flashing message will display while the instrument is warming up. During this
time the icon keys are inactive and you will be unable to start an assay. Go to the
next step when the warm-up message no longer displays.
2.
From the main menu, enter Operator ID and Password, if required. Use
the keypad to enter the Operator ID number at the prompt. Press the Tab key to
advance to the Password prompt to enter the password number.
3.
Press the QC key. Wait for the image of the device to display (Figure 5-2).
Figure 5-2 Start EQC
4.
Remove the EQC device from the storage bay on the right side of the instrument.
Open the cover and using the finger grip, insert the EQC device into the assay
device port until it clicks (Figure 5-3). The instrument will produce two audible
beeps. Close the cover to the assay port.
NOTE: Do not open the cover until the test has completed.