EasyManua.ls Logo

Acteon XMIND DC - Page 11

Acteon XMIND DC
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 11 of 61
ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
CAUTION
MECHANICAL RISK
Before removing the tubehead from the positioning arm, RELEASE THE SPRING.
The sudden opening of the joint may cause damage to people and/or things.
Check that the installation of the unit complies with the mechanical specications of the support (walls, ceiling, etc..) where it is installed.
Adjustments or any kind of attempt of repairing or disassembling must only be performed by qualied and authorized service personnel.
The “Xmind®dc” must not be used in environments or close to environments subjected to mechanical vibrations or mechanical shocks.
CAUTION
ELECTRIC SAFETY
The radiographic system contains high voltage. It is prohibited to inspect internal parts of the system.
Never attempt to open the X-ray tubehead.
The covers on the “Xmind®dc” radiographic system must only be removed by qualied and authorized service personnel.
The unit must be used only in environments that are in compliance with all electrical safety standards set forth for medical environments.
To avoid the risk of electric shock, this device must only be connected to a supply mains with protective earth.
The unit is NOT equipped with protections against penetration of liquids; it will therefore be necessary to make sure that no water or other
liquids penetrate inside in order to avoid short circuits or corrosion.
Always disconnect the radiographic system from the power supply and wait for 2 minutes before beginning to clean, disinfect and maintenance
operations.
Do not connect the X-ray system to a multiple portable socket outlet (MPSO) nor to any type of extension cord.
External equipment intended for connection to signal input, signal output or other connectors shall comply with the relevant product standard
e.g. IEC 60950-1 for IT equipment and the IEC 60601-series for medical electrical equipment. In addition, all such combinations – systems – shall
comply with the safety requirements stated in the general standard CEI EN 60601-1, edition 3, clause 16. Any equipment not complying with the
leakage current requirements in CEI EN 60601-1 shall be kept outside the patient environment i.e. at least 1,5 m from the patient support.
It is mandatory to use an isolation device (Separation Device) to isolate the equipment located outside the patient environment from the equip-
ment located inside the patient environment. In particular such a Separation Device is required when a network or data connection is made. The
requirements on the Separation Device is dened in CEI EN 60601-1, edition 3, clause 16.
For the wall version of “Xmind®dc”:
based on the CEI EN 60601-1, the installation is a permanent type (xed). IT IS NOT ALLOWED TO connect the equipment to the main supply using
a plug.
The cone (beam limiting device) is an APPLIED PART of the system and it is classied as type B.

Related product manuals