ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ANNEX 5
A5. ELECTROMAGNETIC COMPATIBILITY
Electromagnetic compatibility (EMC) is assessed with reference to the following standards:
CEI EN 60601-1-2: 2007, 3rd edition
EMISSION
• CEI EN 55011: 2013
• CEI EN 61000-3-2: 2015
• CEI EN 61000-3-3: 2014
IMMUNITY
• CEI EN 61000-4-2: 2011
• CEI EN 61000-4-3: 2007 + A1: 2008
• CEI EN 61000-4-4: 2013
• CEI EN 61000-4-5: 2007
• CEI EN 61000-4-6: 2011
• CEI EN 61000-4-8: 2013
• CEI EN 61000-4-11:2006
Guidance and manufacturer’s declaration – electromagnetic emissions
“Xmind®dc” is intended to be used in the electromagnetic environment specied below. The customer or the
operator of “Xmind®dc” must ensure that the device is used in this type of environment.
Emission test Conformity Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
“Xmind®dc” uses RF energy only for internal operation.
RF emissions are extremely and attenuated are not likely
to generate interference with electronic equipment in the
vicinity.
RF emissions
CISPR 11
Class B
“Xmind®dc” is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
CEI EN 61000-3-2
Class A
Voltage uctuations/icker
emissions
CEI EN 61000-3-3
Complies
A5A5