7.1. START UP
CAUTION
When all connections are completed, the installer must check the electric safety and functions of the system.
INSTRUCTION
Bring the main switch located on the upper part of the timer to the “I” position (ON)
Bring the key switch to the “I” position (ON)
1. the green light turns on, indicating that the system is powered
2. the LEDs of the set parameters automatically light up
3. the exposure time is shown on the display
CAUTION
If an error is detected when the system is turned on, the anomaly is indicated as follows:
- an intermittent sounds off (beep);
- MALFUNCTIONING INDICATOR LED
intermittently turns on
- the error code (E ….) appears on the display (refer to Chapter 11)
- all control panel functions are inhibited;
In this case turn off the timer and then turn it back on.
If the error should repeat itself, call the “Assistance Service”.
PLEASE NOTE
The exposure time and parameters which appear on the display are the last that were set before the timer was turned off.
If the timer remains inactive for a few minutes, it switches to the stand-by mode.
Press any key on the control panel to restore it to the operative mode.