CHAPTER 10
10.1. CALIBRATION OF THE TUBEHEAD
CAUTION
During this operation there is X-ray output.
It is mandatory to adopt all the safety measures relevant to radioprotection.
INSTRUCTION
1. turn off the timer.
2. turn on the timer by keeping pressed the key.
3. the message TUBE is displayed.
4. select with the key the tubehead to calibrate.
5. take the exposure: on the control button press the X-RAY key and keep it pressed until the acoustic signal
(beep) stops and the yellow LED turns off.
6. once the exposure has been taken, if the display does not show errors, the calibration has been successfully
done.