11.1. ERROR MESSAGES
The following chart gives a list of error messages that may appear while “Xmind®dc” radiographic system is
working.
The chart also includes the causes of the error messages and what to do to solve them.
ERROR
MESSAGES
CAUSE SOLUTION
E00
RX1 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “ASSISTANCE SERVICE”
E01
RX2 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “ASSISTANCE SERVICE”
E02 CORRUPTED EEPROM DATA CALL THE “ASSISTANCE SERVICE”
E03 EEPROM DATA NOT SAVED PROPERLY CALL THE “ASSISTANCE SERVICE”
E07
LINE VOLTAGE VALUE NOT INCLUDED
WITHIN THE NOMINAL VALUE
CALL THE “ASSISTANCE SERVICE”
E08
THE X-ray KEY
ALWAYS SEEMS TO BE PRESSED
MAKE SURE IT IS NOT JAMMED
E09 ANOMALY IN THE CONTROL PANEL CALL THE “ASSISTANCE SERVICE”
E12
THE EXPOSURE
HAS BEEN PREMATURELY INTERRUPTED
KEEP THE X-ray KEY
PRESSED TILL THE END OF THE EXPOSURE
E20 ANOMALY IN THE TRIAC/RELAY CALL THE “ASSISTANCE SERVICE”
E21 ANOMALY IN THE ELECTRONIC CIRCUIT CALL THE “ASSISTANCE SERVICE”
E22 ANOMALY IN THE CONTROL CIRCUIT CALL THE “ASSISTANCE SERVICE”
E23 INCORRECT DIP-SWITCH CONFIGURATION CALL THE “ASSISTANCE SERVICE”
E24
THE CONTROL BUTTON DOES NOT CORRESPOND
TO THE SELECTED TUBEHEAD
CALL THE “ASSISTANCE SERVICE”
E30
THE TUBEHEAD
DOES NOT WORK PROPERLY
CALL THE “ASSISTANCE SERVICE”
E32
THE TUBEHEAD
IS NOT IN THE CORRECT MODE
CALL THE “ASSISTANCE SERVICE”
E33
THE TUBEHEAD
HAS NOT COMPLETED THE EXPOSURE
REPEAT THE EXPOSURE
CALL THE “ASSISTANCE SERVICE”
E40
PROBLEM IN THE FREQUENCY
OR REGULATION
CALL THE “ASSISTANCE SERVICE”
E41
THE TUBEHEAD
IS NOT CALIBRATED
CARRY OUT WITH THE CALIBRATION
CALL THE “ASSISTANCE SERVICE”
E42 EEPROM DATA NOT SAVED PROPERLY CALL THE “ASSISTANCE SERVICE”
E43 CORRUPTED EEPROM DATA CALL THE “ASSISTANCE SERVICE”