18. Replace the damaged guards
19. Clean the old grease of the shaft: should the shaft be damaged, send the pantograph type arm to the
manufacturer for repairs
20. Lubricate the pantograph type arm shaft with lubricating grease (use grease Molikote D) and reposition it in
the bracket
21. Put again the pantograph type arm cable in the bracket and the wall plate, connect it to the terminal board
and put in place the terminal board cover
22. Position the guard slab in the bracket
23. Position the bracket plug
24. Position the plate guard
25. Check the electric contact of the tubehead: if damaged, send the tubehead to the manufacturer for repairs
26. Clean the old grease from the tubehead assembly shaft
27. Grease the assembly shaft of the tubehead with a thin layer of lubricating grease (use grease Molikote D)
28. Position the tubehead again
29. Load the spring of the pantograph type arm B using the key provided
30. Give power and check the correct operation of the radiographic system