Acteon XMIND DC - Page 67

Acteon XMIND DC
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 67 of 61
ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
DEVICE POWER SUPPLY
Type of power supply single phase, alternate
Supply nominal voltage
230 V
115 V
Maximum voltage variation ±10%
Nominal current
5,2A @ 230V
9,5A @ 115V
Supply voltage frequency 50/60 Hz
Maximum line current (measured @ 70 kV, 8 mA, 3,2 s) 9,5 A @ 115 V
Absorbed power
1,2 kVA @ 230V
1,1 kVA @ 115V
Apparent resistance
0,5 Ω
0,2 Ω
Protection fuses (F1 – F2 – F3 – F4)
F 8 A – 250 V
F 12,5 A – 250 V
Circuit protection fuses
(F5) – n° 1 630 mA – 125 V
(F6) – n° 1 500 mA – 125 V
ELECTRICAL CLASSIFICATION (IEC 60601-1)
Protection against electrical shock (insulation class) Class I
Degree of protection against electrical Shock (applied part) Type B (collimator cone)
Protection against harmful ingress of water or particulate matter IP 20
Use with ammable anaesthetics
Not for use in presence of ammable
anaesthetic mixture with air, oxygen or
nitrous oxide.
Sterilization and disinfection methods
The device is supplied not sterile and it must
not be subjected to sterilization
Operation mode
Continuous operation with intermittent X-ray
loading
MECHANICAL DATA
Total weight
19,5 kg (wall mounting)
50 kg (mobile version)
Weight of the tubehead 5,5 kg
Mechanical conguration Wall mounting, top and bottom / Mobile