EasyManuals Logo

Advanced Instruments OsmoPRO MAX User Manual

Advanced Instruments OsmoPRO MAX
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
Chapter 2: Installation and Setup
OsmoPRO MAX Automated Osmometer User Guide
29
Configure Operator/Testing Access, Quality Control, and
System Timeouts (Supervisors only)
Supervisors can perform the following operations from
the Settings > Configuration screen:
Configure permissions to certain functions for
Operator-level users.
Configure log in, sample ID, and replicate
requirements when testing samples.
Configure requirements for quality control testing
frequency using Clinitrol™.
Enable/disable operator login using badge barcode
scanning.
Configure Operator Permissions
To change Operator permissions for functions, complete
the following steps.
1. Press  > Settings > Configuration and login.
2. Select/deselect checkboxes on the Operator
Permissions tab to enable/disable operator access to
the following features:
Set the Language
Set Up Networking Communications
Set the System Date and Time
Running Instrument Calibration
Logging Routine Instrument Cleaning
Change Hardware Options
Using the Controls Log Database
3. Press Save.
4. Press Exit to return to the Settings menu.
Configure Testing Access and Replicates
Authorized users can configure the system to require
users to enter a user ID and/or a sample ID before
testing can start.
1. Press  > Settings >H/W Options and login.
Settings > Configuration > Testing.
2. Select the User ID checkbox to require users to enter
a user ID before testing.
3. Select the Sample ID checkbox to require users to
enter a sample ID before testing.
4. Select the number of replicates Per Sample
(1through10).
5. Select the number of replicates Per Control
(1through10).
6. Press Save.
7. Press Exit to return to the Settings menu.
Configure Quality Control Actions
Supervisors can configure the system to perform one of
the following actions when a sample result is outside the
control limits:
Perform no action
Issue a warning but continue with the test run
Stop the entire test run
They can also configure the system to lock the
instrument if testing with Clinitrol 290 has not been
performed for a certain amount of time.
Configured actions only work with controls that are
entered in the controls database (see Using the Controls
Log Database) and if sample IDs are required during
testing (see Configure Testing Access and Replicates).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoPRO MAX and is the answer not in the manual?

Advanced Instruments OsmoPRO MAX Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoPRO MAX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals