1260 Infinity QuatPump VL User Manual 101
Test Functions and Calibration
8
System Pressure Test
Potential Causes of System Pressure Test Failure
System Pressure Test failed
The test will fail, if the sum of all leaks in the system (pump, autosampler or
column compartment and connections) exceeds the test limit. After isolating
and fixing the cause of the leak, repeat the System Pressure Test to confirm the
system is pressure tight.
Probable cause Suggested actions
1 Purge valve open.
Close the purge valve.
2 Loose or leaky fittings.
Tighten the fitting or exchange the capillary.
3 Pump: Damaged pump seals or pistons.
Run the Leak Rate Test to confirm the leak.
4 Loose purge valve.
Tighten the purge valve nut (14 mm wrench).
5 Autosampler: Loose or leaky fitting.
Tighten or exchange the fitting or capillary.
6 Autosampler: Rotor seal (injection valve).
Exchange the rotor seal.
7 Autosampler: Damaged metering seal or
piston.
Exchange the metering seal. Check the piston
for scratches. Exchange the piston if required.
8 Autosampler: Needle seat.
Exchange the needle seat.
9 Column compartment: Loose or leaky fitting.
Tighten or exchange the fitting or capillary.
10 Column compartment: Rotor seal in optional
valve.
Exchange the rotor seal.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals