102 1260 Infinity QuatPump VL User Manual
8 Test Functions and Calibration
Leak Rate Test
Leak Rate Test
Introduction
The Leak Rate Test is used for verifying the internal tightness of the pump and
helps identifying parts which may have caused a leak.
System requirements
Minimum software revisions:
Lab Advisor B.01.04. SP1 (G1310B Isocratic Pump, G1311B Quaternary
Pump, G4280B Isocratic Pump, G4281B Gradient Pump)
Lab Advisor B.01.04. SP2 (G1311C Quaternary Pump VL, G4220A Binary
Pump, G4220B Binary Pump VL, G5611A Bio-inert Quaternary Pump)
Minimum firmware revision: A.06.34 for G5611A and A.06.33 for all other
pumps.
Test Principle
A solvent can be chosen from available solvent channels and a maximum
target pressure can be defined at which the test will be run. Typically, this is
the maximum pressure specified for the pump. The test can be run with any
solvent compatible to the pump.
Before the test, the pump is flushed with solvent in order to remove air
bubbles, as air bubbles are compressed during the test and therefore would
appear as leaks. Using a degasser is highly recommended.
NOTE
This test does not work in emulation mode. In case of an emulated module, convert to the
original type first.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals