1260 Infinity QuatPump VL User Manual 85
Error Information
7
Module Error Messages
Motor-Drive Power
Error ID: 2041, 2042
The current drawn by the pump motor exceeded the maximum limit.
Blockages in the flow path are usually detected by the pressure sensor in the
damper, which result in the pump switching off when the upper pressure limit
is exceeded. If a blockage occurs before the damper, the pressure increase
cannot be detected by the pressure sensor and the module will continue to
pump. As pressure increases, the pump drive draws more current. When the
current reaches the maximum limit, the module is switched off, and the error
message is generated.
Probable cause Suggested actions
1 Flow path blockage in front of the damper.
Ensure the capillaries and frits between the
pump head and damper inlet are free from
blockage.
2 Blocked passive inlet valve.
Exchange the passive inlet valve.
3 Blocked outlet valve.
Exchange the outlet valve.
4 High friction (partial mechanical blockage) in
the pump drive assembly.
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
5 Defective pump drive assembly.
Please contact your Agilent service
representative.
6 Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals