EasyManuals Logo

Agilent Technologies 1260 Infinity VL User Manual

Agilent Technologies 1260 Infinity VL
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
138 1260 Infinity QuatPump VL User Manual
9 Maintenance
Replacing the Module Firmware
Replacing the Module Firmware
To upgrade/downgrade the module’s firmware carry out the following
steps:
1 Download the required module firmware, the latest LAN/RS-232 FW
Update Tool and the documentation from the Agilent web.
http://www.chem.agilent.com/scripts/cag_firmware.asp.
2 For loading the firmware into the module follow the instructions in the
documentation.
Module Specific Information
There is no specific information for this module.
When The installation of newer firmware might be necessary
if a newer version solves problems of older versions or
to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
to keep all systems on the same (validated) revision or
if a new module with newer firmware is added to a system or
if third part control software requires a special version.
Tools required Description
LAN/RS-232 Firmware Update Tool
OR Agilent Diagnostic Software
OR Instant Pilot G4208A
(only if supported by module)
Parts required # Description
1 Firmware, tools and documentation from Agilent web site
Preparations Read update documentation provided with the Firmware Update Tool.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals