86 1260 Infinity QuatPump VL User Manual
7 Error Information
Module Error Messages
Inlet-Valve Missing
Error ID: 2048, 2052
The active-inlet valve in the module is missing or defective.
The processor checks the presence of the active-inlet valve connector every
2 s. If the connector is not detected by the processor, the error message is
generated.
Temperature Out of Range
Error ID: 2517
The temperature sensor readings in the motor-drive circuit are out of range.
The values supplied to the ADC by the hybrid sensors must be between 0.5 V
and 4.3 V. If the values are outside this range, the error message is generated.
Probable cause Suggested actions
1 Disconnected or defective cable.
Ensure the pins of the active inlet valve
connector are not damaged. Ensure the
connector is seated securely.
2 Disconnected or defective connection cable
(front panel to main board).
Please contact your Agilent service
representative.
3 Defective active inlet valve.
Exchange the active inlet valve.
Probable cause Suggested actions
1 Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals