56 1260 Infinity QuatPump VL User Manual
5 Optimizing Performance
When to use the Seal Wash Function
When to use the Seal Wash Function
Highly concentrated buffer solutions will reduce the lifetime of the seals and
pistons in your pump. The seal wash function allows to maintain the seal
lifetime by flushing the back side of the seal with a wash solvent.
The seal wash function is strongly recommended when buffer concentrations
of 0.1 M or higher will be used for long time periods in the pump.
The active seal wash upgrade can be ordered as G1398A.
The seal wash function comprises a support ring, secondary seal, gasket and
seal holder for both piston sides. Place a wash bottle filled with 90 % water /
10 % isopropanol above the pump in the solvent cabinet. The peristaltic pump
moves a flow through the pump head removing all possible buffer crystals
from the back of the pump seal. This mixture prevents growth of algae or
bacteria in the wash bottle and reduces the surface tension of the water.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals