88 1260 Infinity QuatPump VL User Manual
7 Error Information
Module Error Messages
Servo Restart Failed
Error ID: 2201, 2211
The pump motor in the module was unable to move into the correct position
for restarting.
When the module is switched on, the first step is to switch on the C phase of
the variable reluctance motor. The rotor should move to one of the C positions.
The C position is required for the servo to be able to take control of the phase
sequencing with the commutator. If the rotor is unable to move, or if the C
position cannot be reached, the error message is generated.
Probable cause Suggested actions
1 Disconnected or defective cable.
Please contact your Agilent service
representative.
2 Blocked passive inlet valve.
Exchange the passive inlet valve.
3 Mechanical blockage of the module.
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
4 Defective pump drive assembly.
Please contact your Agilent service
representative.
5 Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals