EasyManua.ls Logo

Agilent Technologies 1260 Infinity VL - Module Error Messages; Solvent Zero Counter

Agilent Technologies 1260 Infinity VL
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1260 Infinity QuatPump VL User Manual 79
Error Information
7
Module Error Messages
Module Error Messages
These errors are pump specific.
Solvent Zero Counter
Error ID: 2055, 2524
Pump firmware version A.02.32 and higher allow to set solvent bottle fillings
in the data system. If the volume level in the bottle falls below the specified
value the error message appears when the feature is configured accordingly.
Probable cause Suggested actions
1 Volume in bottle below specified volume.
Refill bottles and reset solvent counters.
2 Incorrect setting.
Make sure the limits are set correctly.

Table of Contents

Related product manuals