1260 Infinity QuatPump VL User Manual 121
Maintenance
9
Maintenance of a Pump Head Without Seal Wash Option
Maintenance of a Pump Head Without Seal Wash Option
For a complete list of parts see “Pump Head Assembly Without Seal Wash” on
page 140.
When In case of maintenance or pump head internal leaks.
Tools required Description
Wrench 1/4 inch
Hexagonal key, 4 mm
Parts required # p/n Description
1 01018-23702 Insert tool
1 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default
OR 1 0905-1420 PE seals (pack of 2)
1 5063-6586 Sapphire piston
Preparations Switch off pump at the main power switch
Remove the front cover
“Removing the Pump Head Assembly” on page 119