92 1260 Infinity QuatPump VL User Manual
7 Error Information
Module Error Messages
Wait Timeout
Error ID: 2053
When running certain tests in the diagnostics mode or other special
applications, the pump must wait for the pistons to reach a specific position,
or must wait for a certain pressure or flow to be reached. Each action or state
must be completed within the timeout period, otherwise the error message is
generated.
Possible Reasons for a Wait Timeout:
· Pressure not reached.
· Pump channel A did not reach the delivery phase.
· Pump channel B did not reach the delivery phase.
· Pump channel A did not reach the take-in phase.
· Pump channel B did not reach the take-in phase.
· Solvent volume not delivered within the specified time.
Probable cause Suggested actions
1 Purge valve open.
Ensure that purge valve is closed.
2 Leak at fittings, purge valve, active inlet
valve, outlet valve or piston seals.
Ensure pump components are seated correctly.
If there are still signs of a leak, exchange the
appropriate seal (purge valve, active inlet valve,
outlet valve, piston seal).
3 Flow changed after starting test.
Ensure correct operating condition for the
special application in use.
4 Defective pump drive assembly.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals