1260 Infinity QuatPump VL User Manual 105
Test Functions and Calibration
8
Leak Rate Test
Potential Causes of Leak Rate Test Failure
Secondary Leak
If a leak is found for movement of piston 2 (secondary leak), the following
reasons are possible:
Primary Leak
If a leak is found for movement of piston 1 (primary leak), any leak described
for piston movement 2 will cause a failure for piston 1 as well, as the liquid
can move through the outlet valve to chamber 2. Such cases need to be
identified as described before. Additionally, following causes are possible:
Internal Outlet Valve Leak
A leak of the outlet valve will be identified separately (internal outlet valve
leak) by calculating the difference between leak rate 1 and leak rate 2. If the
second leak rate is higher than the first one, this is due to a flow back through
the outlet valve.
Probable cause Suggested actions
1 System not flushed properly
Flush system for several minutes
2 Degassing efficiency is low
Check degasser performance
3 Purge valve not closed or defect
Check purge valve
4 Blank nut not installed tightly
Tighten or replace blank nut
5 Outlet valve leaking (read below)
Replace outlet valve
6 Leak at piston 2 or seal in chamber 2
Inspect piston, replace piston and/or seal
Probable cause Suggested actions
1 Leak at piston 1 or seal in chamber 1
Inspect piston, replace piston and/or seal
2 Leak at inlet valve
Replace inlet valve or inlet valve cartridge (AIV only)
Probable cause Suggested actions
1 Leak at outlet valve
Replace the part which has failed and re-run the test.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals