78 1260 Infinity QuatPump VL User Manual
7 Error Information
General Error Messages
Open Cover
Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed, the fan is switched off, and the error message is generated.
Probable cause Suggested actions
1 The top foam was removed during operation.
Please contact your Agilent service
representative.
2 Foam not activating the sensor.
Please contact your Agilent service
representative.
3 Dirty or defective sensor.
Please contact your Agilent service
representative.
4 Rear of the module is exposed to strong
direct sunlight.
Ensure that the rear of module is not directly
exposed to strong sunlight.