1260 Infinity QuatPump VL User Manual 93
Error Information
7
Module Error Messages
Degasser: cannot read signal
Error ID: 2243
The pump board gets no or wrong pressure signals from the built-in degasser.
Degasser: limit not reached
Error ID: 2244
This error is thrown, if the degasser does not become ready after 8 min, i.e. is
higher than 180 mbar.
Probable cause Suggested actions
1 Degasser board defect, missing or not
connected to the pump main board.
Please contact your Agilent service
representative.
2 Degasser sensor defect or not connected to
degasser board
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Liquid in degasser tubing.
Please contact your Agilent service
representative.
2 Leak in degasser tubing or chamber.
Please contact your Agilent service
representative.
3 Degasser vacuum pump defect.
Please contact your Agilent service
representative.