40 1260 Infinity QuatPump VL User Manual
3 Installing the Pump
Flow Connections of the Pump
1 Remove the front cover by pressing the snap fasteners on both sides.
Figure 7 Removing the Front Corver
2 Place the solvent cabinet on top of the quaternary pump.
3 Put the bottle-head assemblies into empty solvent reservoirs and place the
bottle in the solvent cabinet.
4 Connect the inlet tubes from the bottle-head assemblies to the inlet
connectors A to D at the right hand side of the vacuum degasser, see
Figure 8 on page 41. Fix the tubes in the tube clips of the pump.
5 Connect the solvent tubes from the MCGV inlet to the outlets of the vacuum
degasser.
6 Using a piece of sanding paper connect the waste tubing to the purge valve
and place it into your waste system.
7 If the pump is not part of an Agilent 1260 Infinity system stack or placed on
the bottom of a stack, connect the waste tube to the waste outlet of the
pump leak handling system.
8 Connect the pump outlet capillary (pump to injection device) to the outlet
of the purge valve.
9 Fill solvent reservoirs with your mobile phase.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity VL and is the answer not in the manual?

Agilent Technologies 1260 Infinity VL Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity VL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals