3-40 Agilent 4155B/4156B Users Guide Vol.2, Edition 5
Measurement Functions
Trigger Function
Setup and restrictions
You cannot perform trigger outputs together with trigger inputs. You must select
either trigger output or trigger input.
To use a trigger function, you must enable the trigger function and select either
TRIG OUT or TRIG IN in the TRIGGER SETUP table on the MEASURE:
OUTPUT SEQUENCE screen. Then the trigger inputs or outputs are performed
automatically after you start a measurement by selecting a measurement
front-panel key (Single, Repeat, or Append).
The trigger output function is not available for sampling measurements.
When you perform knob sweep measurements, the trigger function is not
available.
For the electrical specifications of trigger signals, refer to User's Guide General
Information.