Agilent 4155B/4156B Users Guide Vol.2, Edition 5 1-5
Measurement Units
Source/Monitor Unit (SMU)
Three types of SMUs are available:
HRSMU (high resolution SMU)
Force and measure: up to ±100 V or ±100 mA.
Maximum output power: 2 W.
Minimum current measurement range: 10 pA with 1 fA resolution.
Only the 4156B has HRSMUs. The 4156B has four HRSMUs.
MPSMU (medium power SMU)
Force and measure: up to ±100 V or ±100 mA.
Maximum output power: 2 W.
The 4155B has four MPSMUs, and the 41501A/B can be equipped with
either two MPSMUs or one HPSMU.
HPSMU (high power SMU)
Force and measure: up to ±200 V or ±1 A.
Maximum output power: 20 W.
Only the 41501A/B has HPSMU. The 41501A/B can be equipped with
either two MPSMUs or one HPSMU.
HPSMUs and HRSMUs can be connected to test devices by Kelvin connection.
Each SMU has a compliance feature that limits output voltage or current to prevent
damage to your devices. When the SMU forces voltage, you can specify I
compliance. When the SMU forces current, you can specify V compliance.
For details about the compliance setting range and resolution, see Compliance in
Chapter 3.
The following figures and tables show the output and measurement ranges of each
SMU type.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156b and is the answer not in the manual?

Agilent Technologies 4156b Specifications

General IconGeneral
BrandAgilent Technologies
Model4156b
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals