Agilent 4155B/4156B User’s Guide Vol.2, Edition 5 8-15
If You Have A Problem
If Errors Occur
1xx15 Measurement units that are not supported are detected.
1xx19 Emergency occurred but the cause is unknown. This is displayed, for
example, when unit is known but cause is unknown.
1xx90 AD converter test reached timeout during calibration or self-test for a
measurement unit.
1xx91 FIFO (first-in, first-out) for AD converter overflowed because SMU
controller takes long time to read measurement data.
1xx92 Calibration or diagnostics was aborted by an emergency or *RST
command.
1xx94 The 4155B/4156B was turned on before the 41501A/B.
1xx97 Communication failed between HOST controller and SMU controller.
Or calibration/diagnostics was performed, but HOST controller
couldn't receive the result from SMU controller.
10030 VSUs and VMUs failed default test of calibration.
10031 VSUs and VMUs failed function check.
10032 VSUs failed gain or offset calibration.
10033 VMUs failed gain or offset calibration.
10034 VMUs failed differential mode 2 V range gain or offset calibration.
10035 VSUs failed gain and offset calibration, VMUs failed gain and offset
calibration, or VMU failed differential mode 2 V range gain and offset
calibration.
10036 VMUs failed differential mode 0.2 V range gain or offset measurement.
10037 VMUs failed differential mode 0.2 V range gain and offset calibration.
10038 VMUs and VSUs failed CMR (Common Mode Rejection) amp
adjustment.
10040 VSU1 and VMU1 failed ±20 V measurement self-test in 20 V range.
10041 VSU2 and VMU2 failed ±20 V measurement self-test in 20 V range.
10042 VSU1 and VMU2 failed ±20 V measurement self-test in 20 V range.
10043 VMU2 and VMU1 failed ±20 V measurement self-test in 20 V range.
10044 VSU1 and VMU1 failed ±2 V measurement self-test in 2 V range.
10045 VSU2 and VMU2 failed ±2 V measurement self-test in 2 V range.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156b and is the answer not in the manual?

Agilent Technologies 4156b Specifications

General IconGeneral
BrandAgilent Technologies
Model4156b
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals