3-24 Agilent 4155B/4156B Users Guide Vol.2, Edition 5
Measurement Functions
Knob Sweep Function
VAR2
POINTS
5 Sets Number of steps for VAR2 channel. This softkey displays present
setting.
setting
Press this softkey to display the VAR2 number of steps in the data
entry area. Then, you can change the setting value by using the
rotary knob, numeric keys, or arrow keys in the Edit key group.
setting range : 1 to 128
default
VAR2 number of steps on MEASURE: SWEEP SETUP screen
COMPLI-
ANCE
2.00 V Sets compliance value for VAR2 channel. This softkey displays present
setting.
setting
Press this softkey to display the VAR2 compliance value in the data
entry area. Then, you can change the setting value by using the
rotary knob, numeric keys, or arrow keys in the Edit key group.
setting range
Setting range depends on the measurement unit. See Chapter 1.
default
VAR2 compliance value on MEASURE: SWEEP SETUP screen