Copyright 2007 Agilent Technologies
Change 14
Change the Measurement Uncertainty value of 1 m Impedance Measurement Accuracy Test
for 1 m DC Resistance Measurement Accuracy Test (Opt. 001 Only) (Page 9-43) to the
following information.
Standard
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm ±1.05 mohm mohm ± 0.030 mohm
100 kohm ±0.87 kohm kohm ± 0.014 kohm
Change 14
Change the Measurement Uncertainty value of 2 m Impedance Measurement Accuracy Test
for 2 m Capacitance Measurement Accuracy Test (Page 9-44) to the following information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
100 pF 1 kHz Cp ±0.19 pF ± 0.011 pF
D ±0.0019 ± 0.000073
100 pF 10 kHz Cp ±0.51 pF ± 0.013 pF
D ±0.0051 ± 0.00014
100 pF 20 kHz
2
Cp ±1.10 pF ± 0.022 pF
D ±0.0110 ± 0.000094
1 nF 100 Hz Cp ±0.002 nF ± 0.000090 nF
D ±0.0020 ± 0.000075
1 μF 100 Hz Cp ±0.0018 μF ± 0.000077 μF
D ±0.0018 ± 0.000047
1 μF 1 kHz Cp ±0.0011 μF ± 0.000069 μF
D ±0.0011 ± 0.000033
1 μF 10 kHz Cp ±0.0026 μF ± 0.00014 μF
D ±0.0026 ± 0.000062
Change 15
Change the Measurement Uncertainty value of 2 m Impedance Measurement Accuracy Test
for 2 m Resistance Measurement Accuracy Test (Page 9-44) to the following information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm 100 Hz R ±0.68 mohm mohm ± 0.030 mohm
100 mohm 1 kHz R ±0.72 mohm mohm ± 0.034 mohm