4263B
SOURce
Subsystem
COMMAND
P
ARAMETER
SOURce
:FREQuency
[
:CW
]
<
numeric
value
>
[
HZ
j
KHZ
]
:VOLTage
[
:LEVel
]
[
:IMMediate
]
[:AMPLitude]
<
numeric
value
>
[
MV
j
V
]
:OFFSet
<
numeric
value
>
[
MV
j
V
]
:SOURce
f
INTernal
j
EXTernal
g
:STATe
<
Boolean
>
:SOURce:FREQuency[:CW]
<
numeric
value
>
[HZ
j
KHZ]
Sets
or
queries
the
test
signal
frequency
.
<
numeric
value
>
is
,
numeric
100, 120,
1000 (=
1k),
10000
(=
10
k),
20000
(=
20
k;
option
002
only),
100000
(= 100
k) (Hz)
MINimum
100 (Hz)
MAXimum
100000
(=
100 k)
(Hz)
Query
response
is
a
numeric
value
in
<
NR3
>
format.
:SOURce:V
OL
T
age[:LEV
el][:IMMediate][:AMPLitude]
<
numeric
value
>
[MV
j
V]
Sets
or
queries
the
test
signal
level.
<
numeric
value
>
is
,
numeric
20
to
1000
(mV),
(5
mV
step)
MINimum
20
(mV)
MAXimum
1
(V)
Query
response
is
a
numeric
value
in
<
NR3
>
format.
:SOURce:V
OL
T
age[:LEV
el][:IMMediate]:OFFSet
<
numeric
value
>
[MV
j
V]
Sets
or
queries
the
DC
bias
voltage
.
<
numeric
value
>
is
,
numeric
0, 1.5, 2 (V)
MINimum
0(V)
MAXimum
2(V)
Query response is a numeric value in
<
NR3
>
format.
:SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet:SOURce
f
INTernal
j
EXTernal
g
Sets or queries the DC bias voltage source: Internal
or External.
INTernal
Internal voltage source.
EXTernal
External voltage source.
Query response is
INT
or
EXT
.
GPIB Reference 5-29

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals