4263B
T
o
Retrieve
Data
Eciently
The
basic
procedure
to
retrieve
measurement
data
is
described
in
\To
Trigger a
Measurement".
This
section
describes
how
to
retrieve
the measurement
data eciently
.
T
o
Use
a
Data
Buer
Y
ou
can
use
the
data
buer
to
temporarily
store
the results
of several
measurements and
then
later
read
out
these
results
in
batch.
The
following
commands
use
the
data
buer function:
:DATA?
:DATA:POIN
:DATA:FEED
:DATA:FEED:CONT
F
or
example
,
:
OPTION
BASE
1
DIM
D(600)
:
OUTPUT
717;":TRIG:SOUR
BUS"
!
OUTPUT
717;":DATA:POIN
BUF1,200"
OUTPUT
717;":DATA:FEED
BUF1,'CALC1'"
OUTPUT
717;":DATA:FEED:CONT
BUF1,ALW"
!
FOR
I=1
TO
200
OUTPUT
717;":TRIG"
NEXT
I
!
OUTPUT
717;":DATA?
BUF1"
ENTER
717;D(*)
PRINT
D(*)
:
Remote Operation (To Control from a Computer) 4-25