4263B
T
est
Signal
Level
A
ccuracy
T
est
The
4263B
's
test
signal
level
is
measured
with
an
AC
voltmeter.
Specication
T
est
Signal
Level
A
ccuracy:
6
(10
%+
10 mV)
T
est
Equipment
Description
Recommended Model
Multimeter
3458A
61
cm BNC(m)-BNC(m)
Cable p/n
8120-1839
Dual
Banana-BNC(f)
A
dapter
p/n
1251-2277
Procedure
1.
Reset
the
4263B
.
2.
Set up
the
equipment
as
shown
in
Figure
9-2
.
Figure 9-2. T
est Signal Level A
ccuracy T
est Setup
3. Set the 3458A Multimeter to the Synchronously Sub-sample A
C voltage measurement mode
using the following procedure:
a. Press
4
ACV
5
to set the measurement mode to A
C voltage
.
b. Press
4
Shift
54
S
54
9
54
9
54
9
5
to display
SETACV
.
c. Press
4
7
54
9
54
9
54
9
5
to display
SYNC
, then press
4
Enter
5
.
4. Set the test signal level to 50 mV using
.
Maintenance 9-7

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals