Agilent Technologies 4263B - Transformer Measurement Functional Test

Agilent Technologies 4263B
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4263B
Transformer
Measurement
Functional
T
est
(Opt.
001
Only)
The
4263B
's
transformer
measurement
function
is
tested
using
a
power splitter
.The
4263B
measures
a
transformer's
turn
ratio
by
stimulating a
transformer and
comparing the
primary
voltage
to
the
secondary
voltage
.
In this
test, the
4263B stimulates
the power
splitter,
instead
of
the
transformer
,
and
compares
the two
voltages across
the power
splitter.
T
est
Equipment
Description
Recommended Model
Transformer
Test
Fixture 16060A
30
cm BNC(m)-BNC(m)
Cable
p/n
8120-1838,
3
ea.
Procedure
1.
Reset
the
4263B
.
2.
Set
up
the
equipment
as
shown in
Figure 9-9
.
Figure
9-9.
Transformer Measurement
Functional
T
est
Setup
3. Set the 4263B measurement conditions as follows:
Measurement P
arameter:
L2-N
Measurement Time Mode:
LONG
Test Signal Level:
1 V (Default Setting)
Test Signal Frequency:
1 kHz (Default Setting)
Trigger Mode:
Man
4. Set the transformer test xture's A:B switch to the 1:N position.
Maintenance 9-47