Agilent Technologies 4263B - Measurement Range; Measurement Accuracy

Agilent Technologies 4263B
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4263B
T
est Cable
Lengths
0
m, 1
m, 2
m, 4
m(@
100 Hz,
120 Hz,
1
kHz)
0
m, 1
m, 2
m(@
10 kHz,
20
kHz)
0
m, 1
m(@
100 kHz)
Measurement
Time Mode
Short,
Medium, and
Long
Ranging
A
uto and
Hold
(manual)
A
veraging
1
to 256
Trigger
Mode
Internal,
Manual,
External,
and
Bus
Trigger
Delay
Time
0
to
9.999
seconds
in
0.001
seconds
steps
Measurement
Range
P
arameter
Measurement
Range
Z,
R,
X
1
m
to
100
M
Y
,
G,
B
10
nS
to
1000
S
C 1
pF
to 1
F
L 10
nH
to
100
kH
P
arameter
Measurement
Range
D 0.0001
to
9.9999
Q 0.1
to
9999.9
6
(
)
0
180
to
180
Rdc 1
m
to
100
M
Measurement
A
ccuracy
The
following
conditions
must
be met:
1.
W
arm
up
time:
15 min
2. OPEN
and SHORT
corrections have
been
performed.
3. T
est Signal
Level V
osc
=50
mV,
100 mV
,
250
mV
,
500
mV
,
1
V
(Measurement accuracy
at V
osc
other than
above listed
points is
typical data.)
4. Ambient temperature: 23
6
5
C
Within
the temperature (T) range of 0 to 45
C,
multiply the basic accuracy by the following
temperature
induced error
,
8
C
T
<
18
C, or
28
C
<
T
38
C:
2
2
0
C
T
<
8
C, or 38
C
<
T
45
C:
2
4
Specications 8-3