EasyManua.ls Logo

Agilent Technologies 4263B - Page 15

Agilent Technologies 4263B
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C
Copyright 2007 Agilent Technologies
Change 19
Change the Measurement Uncertainty value of 4 m Impedance Measurement Accuracy Test
for 4 m DC Resistance Measurement Accuracy Test (Opt. 001.Only) (Page 9-45) to the
following information.
Standard
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm ±1.29 mohm mohm ± 0.030 mohm
100 kohm ±0.87 kohm kohm ± 0.013 kohm