Page #172 background imageLoading...
Page #172 background image
4263B
:D
AT
A[:DA
T
A]?
f
BUF1
j
BUF2
g
Returns
the data
in
a
data
buer
,
BUF1
or
BUF2
,
according
to
the format
determined by
the
FORMat
subsystem commands
.
(query
only)
BUF1
and
BUF2
are
used
to
store
up
to
200
measurement
results
in
each.
The size
of the
data
buer
is specied
by
the
:DATA:POINts
command.
This
query
needs
parameter
BUF1
or
BUF2
,
which
is
the
name
of
the
data buer
to be
read. So
,
the
query
form
is
:DATA[:DATA]?
BUF1
or
:DATA[:DATA]?
BUF2
.
Reading
the
data
buer
returns
its
pointer
to
the
head
of
the
buer
.
After that,
data will
be
stored
from
the
head
of
the
buer
.
Query
response
is
<
set1
>
,
<
set2
>
,.
..
<
setn
>
Where
,
<
set1
>
:
Data
set
of
the
rst
measured
point
<
set2
>
:
Data
set
of
the
second
measured
point
.
.
.
<
setn
>
:
Data
set
of
the
last
measured
point
Every
data
sets
contains
the
following
data:
<
stat
>
,
<
data
>
,
<
comp
>
Where
,
<
stat
>
Measurement
status
0
:
Normal
1
:
Overload
2
:
No-contact
<
data
>
Measurement
data
<
comp
>
Comparison
result
0
:
Comparator
OFF
1
:
In
2
:
High
4
:
Low
8:
No-contact
:D
A
TA[:D
AT
A]?
f
IMON
j
VMON
g
Reads
the
level monitor
value.
(query only)
IMON Reads
the current
monitor.
VMON Reads
the
voltage
monitor
.
Query response is a numeric value in
<
NR3
>
format.
GPIB Reference 5-17

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals