4263B
T
able 9-9.
0m
Resistance
Measurement
T
est
(LONG)
Settings
Standard Test
Signal
Frequency
100
m
100
Hz
1
kHz
0
m
Resistance
Measurement
A
ccuracy
T
est
(Meas
.
Time
Mode:
MED).
40.
Set
the
4263B
measurement
conditions
as
follows:
Measurement
P
arameter:
R-X
Measurement
Time Mode:
MED
T
est
Signal
Level:
500
mV
T
est
Signal
Frequency:
100
Hz
Trigger
Mode:
Man
41.
Connect
the
100
m
Standard
Resistor
to
the
4263B
's
UNKNOWN
terminals
.
42.
Press
to
start
the
measurement.
43.
Record
the
4263B
readings
of
R
on
the
calculation
sheet.
44.
Calculate
the
test
results
according
to
the
calculation
sheet,
and
record
the
result
into
the
performance
test
record.
T
able
9-10.
0
m
Resistance
Measurement
T
est
(MED)
Settings
Standard T
est
Signal
Frequency
100
m
100
Hz
0
m
Resistance
Measurement
A
ccuracy
T
est
(Meas
.
Time
Mode:
SHORT).
45.
Set
the
4263B
measurement
conditions as
follows:
Measurement
P
arameter:
R-X
Measurement Time Mode:
SHORT
Test Signal Level:
500 mV
Test Signal Frequency:
100 Hz
Trigger Mode:
Man
46.
Connect the 100 m Standard Resistor to the 4263B's UNKNOWN terminals
.
47. Press
to start the measurement.
48. Record the 4263B readings of R on the calculation sheet.
49. Calculate the test results according to the calculation sheet, and record the result into the
performance test record.
9-16 Maintenance

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals