EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>4263B

Agilent Technologies 4263B User Manual

Agilent Technologies 4263B
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #291 background imageLoading...
Page #291 background image
4263B
2
m Impedance
Measurement A
ccuracy
T
est
2
m Capacitance
Measurement
A
ccuracy
T
est
Standard T
est Signal
Frequency
Parameter
Measured
Test
Limits T
est
Result
1
Measurement
Uncertainty
100
pF
1
kHz Cp
6
0.19
pF
pF
6
0.018
pF
D
6
0.0019
6
0.0001
100
pF
10
kHz
Cp
6
0.51
pF
pF
6
0.029
pF
D
6
0.0051
6
0.0002
100
pF
20
kHz
2
Cp
6
1.10
pF
pF
6
0.029
pF
D
6
0.0110
6
0.0002
1000
pF 100
Hz
Cp
6
2.0
pF
pF
6
0.18
pF
D
6
0.0020
6
0.0001
1

F
100
Hz
Cp
6
0.0018

F

F
6
0.00044

F
D
6
0.0018
6
0.00015
1

F
1
kHz
Cp
6
0.0011

F

F
6
0.00020

F
D
6
0.0011
6
0.0001
1

F
10
kHz
Cp
6
0.0026

F

F
6
0.00044

F
D
6
0.0026
6
0.0004
1
T
est
Result
= Measured
V
alue
0
Standard's
Calibration
V
alue
2
Option
002
only
.
2
m
Resistance
Measurement
A
ccuracy T
est
Standard
T
est
Signal
Frequency
P
arameter
Measured
T
est
Limits
T
est
Result
1
Measurement
Uncertainty
100 m
100 Hz
R
6
0.68
m
m
6
0.10 m
100
m
1
kHz
R
6
0.72
m
m
6
0.10
m
1
T
est
Result
=
Measured
V
alue
0
Standard's
Calibration
V
alue
2
m
DC
Resistance
Measurement
A
ccuracy
T
est
(Opt.
001
Only)
Standard
T
est
Limits
T
est
Result
1
Measurement
Uncertainty
100 m
6
1.13 m
m
6
0.10 m
100 k
6
0.87 k
k
6
0.021 k
1
Test Result = Measured V
alue
0
Standard's Calibration V
alue
9-44 Maintenance Record-6

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals