4263B
2
m Impedance
Measurement A
ccuracy
T
est
2
m Capacitance
Measurement
A
ccuracy
T
est
Standard T
est Signal
Frequency
Parameter
Measured
Test
Limits T
est
Result
1
Measurement
Uncertainty
100
pF
1
kHz Cp
6
0.19
pF
pF
6
0.018
pF
D
6
0.0019
6
0.0001
100
pF
10
kHz
Cp
6
0.51
pF
pF
6
0.029
pF
D
6
0.0051
6
0.0002
100
pF
20
kHz
2
Cp
6
1.10
pF
pF
6
0.029
pF
D
6
0.0110
6
0.0002
1000
pF 100
Hz
Cp
6
2.0
pF
pF
6
0.18
pF
D
6
0.0020
6
0.0001
1
F
100
Hz
Cp
6
0.0018
F
F
6
0.00044
F
D
6
0.0018
6
0.00015
1
F
1
kHz
Cp
6
0.0011
F
F
6
0.00020
F
D
6
0.0011
6
0.0001
1
F
10
kHz
Cp
6
0.0026
F
F
6
0.00044
F
D
6
0.0026
6
0.0004
1
T
est
Result
= Measured
V
alue
0
Standard's
Calibration
V
alue
2
Option
002
only
.
2
m
Resistance
Measurement
A
ccuracy T
est
Standard
T
est
Signal
Frequency
P
arameter
Measured
T
est
Limits
T
est
Result
1
Measurement
Uncertainty
100 m
100 Hz
R
6
0.68
m
m
6
0.10 m
100
m
1
kHz
R
6
0.72
m
m
6
0.10
m
1
T
est
Result
=
Measured
V
alue
0
Standard's
Calibration
V
alue
2
m
DC
Resistance
Measurement
A
ccuracy
T
est
(Opt.
001
Only)
Standard
T
est
Limits
T
est
Result
1
Measurement
Uncertainty
100 m
6
1.13 m
m
6
0.10 m
100 k
6
0.87 k
k
6
0.021 k
1
Test Result = Measured V
alue
0
Standard's Calibration V
alue
9-44 Maintenance Record-6