5.
Select
OpenMeas
and press
.
The OPEN
correction is
performed.
6.
Set
the
measurement
parameter
to
R-X.
7.
Connect
the
10
k
standard
resistor
instead
of
the
OPEN
termination.
8.
Press
to
hold
the
measurement
range
.
9.
Press to
activate
the
contact
check
function.
10.
Conrm that
the 4263B
displays
the
measurement
values
.
11.
Remove the
test leads
BNC
connector
from
the
Lpot
terminal
of
the
standard
resistor
.
12.
Conrm that
the 4263B
displays
N.C.
(No
Contact).