Change 5
Add TAR in 1 m Impedance Measurement Accuracy Test (Page 9-43) as follows.
Measurement:
Capacitance
Cable length:
1 m
Measurement time:
Long
Test signal level:
1 V
Status
PASS
Standard
Frequency [Hz]
Parameter Limits Result
1
Uncertainty
TAR
100 pF 1 k Cp ±0.19 pF
pF
± 0.0077 pF
24.59
100 pF 1 k D ±0.0019 ± 0.000063
30.52
100 pF 10 k Cp ±0.48 pF
pF
± 0.012 pF
40.75
100 pF 10 k D ±0.0048 ± 0.00013
37.17
100 pF 20 k Cp ±1.10 pF
pF
± 0.024 pF
47.09
100 pF 20 k D ±0.0110 ± 0.000085
130.36
100 pF 100 k Cp ±1.37 pF
pF
± 0.010 pF
136.75
100 pF 100 k D ±0.0137 ± 0.00016
87.72
1 nF 100 Cp ±0.002 nF
nF
± 0.000091 nF
21.61
1 nF 100 D ±0.0020 ± 0.000063
32.10
1 μF 100 Cp ±0.0018 μF
μF
± 0.000075 μF
24.28
1 μF 100 D ±0.0018 ± 0.000047
38.74
1 μF 1 k Cp ±0.0011 μF
μF
± 0.000069 μF
15.58
1 μF 1 k D ±0.0011 ± 0.000033
33.55
1 μF 10 k Cp ±0.0026 μF
μF
± 0.00014 μF
19.09
1 μF 10 k D ±0.0026 ± 0.000061
43.04
1 μF 100 k Cp ±0.0207 μF
μF
± 0.00016 μF
132.97
1 μF 100 k D ±0.0207 ± 0.00029
73.20
Note 1: Result = Measurement – Calibrated Value of Standard
Measurement:
Resistance
Cable length:
1 m
Measurement time:
Long
Test signal level:
1 V
Status
PASS
Standard
Frequency [Hz]
Parameter Limits Result
1
Uncertainty
TAR
100 m
100
R ±0.60 m m ± 0.030 m
20.44
100 m
1k
R ±0.60 m m ± 0.031 m
19.33
Note 1: Result = Measurement – Calibrated Value of Standard
Measurement:
DC Resistance
Cable length:
1 m
Measurement time:
Long
Test signal level:
1 V
Status
PASS
Standard
Frequency [Hz]
Limits Result
1
Uncertainty TAR
100 m R ±1.05 m m
± 0.030
m
35.88
100 k R ±0.87 k m ± 0.014 k
26.19
Note 1: Result = Measurement – Calibrated Value of Standard